Show simple item record

2014-04-01Zeitschriftenartikel DOI: 10.1097/MBP.0000000000000037
Disagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values.
dc.contributor.authorAdler, Carolin
dc.contributor.authorEllert, Ute
dc.contributor.authorNeuhauser, Hannelore
dc.date.accessioned2018-05-07T18:45:59Z
dc.date.available2018-05-07T18:45:59Z
dc.date.created2015-12-23
dc.date.issued2014-04-01none
dc.identifier.otherhttp://edoc.rki.de/oa/articles/reXns3BKb8ZM/PDF/23fPFIQ2gfHg.pdf
dc.identifier.urihttp://edoc.rki.de/176904/2230
dc.description.abstractObjective(s): Oscillometric blood pressure (BP) measurement devices frequently replace the standard mercury sphygmomanometer. Comparisons of oscillometric devices are rare, but their agreement is important to ensure comparability of BP data. This study aims to compare two oscillometric devices, Datascope Accutorr Plus and Omron HEM-705CP II, and to develop BP conversion models. Methods: A sample of 109 adults aged 21–64 years were subjected to alternate same-arm BP measurements according to the International Protocol revision 2010 for the validation of BP measuring devices in adults of the European Society of Hypertension. Results: A total of 327 BP measurement pairs were obtained. Datascope systolic blood pressure (SBP) pairs, in mmHg, were optimal (eng
dc.language.isoeng
dc.publisherRobert Koch-Institut, Epidemiologie und Gesundheitsberichterstattung
dc.subjectHumanseng
dc.subjectFemaleeng
dc.subjectMaleeng
dc.subjectReproducibility of Resultseng
dc.subjectAdulteng
dc.subjectMiddle Agedeng
dc.subjectYoung Adulteng
dc.subjectBlood Pressureeng
dc.subjectBlood Pressure Determination/instrumentationeng
dc.subjectHypertension/diagnosiseng
dc.subjectHypertension/epidemiologyeng
dc.subjectOscillometry/instrumentationeng
dc.subject.ddc610 Medizin
dc.titleDisagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values.
dc.typeperiodicalPart
dc.identifier.urnurn:nbn:de:0257-10042556
dc.identifier.doi10.1097/MBP.0000000000000037
dc.identifier.doihttp://dx.doi.org/10.25646/2155
local.edoc.container-titleBlood Pressure Monitoring
local.edoc.container-textThis is a non-final version of an article published in final form in Blood Pressure Monitoring 2014(19)2, pp. 109-117.
local.edoc.fp-subtypeArtikel
local.edoc.type-nameZeitschriftenartikel
local.edoc.container-typeperiodical
local.edoc.container-type-nameZeitschrift
local.edoc.container-urlhttp://journals.lww.com/bpmonitoring/Abstract/2014/04000/Disagreement_of_the_two_oscillometric_blood.9.aspx
local.edoc.container-publisher-nameLippincott, Williams & Wilkins
local.edoc.container-volume19
local.edoc.container-issue2
local.edoc.container-year2014

Show simple item record