Calibration of blood pressure data after replacement of the standard mercury sphygmomanometer by an oscillometric device and concurrent change of cuffs
dc.contributor.authorNeuhauser, Hannelore
dc.contributor.authorEllert, Ute
dc.contributor.authorThamm, Michael
dc.contributor.authorAdler, Carolin
dc.date.accessioned2018-05-07T18:46:09Z
dc.date.available2018-05-07T18:46:09Z
dc.date.created2015-12-23
dc.date.issued2014-08-20none
dc.identifier.otherhttp://edoc.rki.de/oa/articles/re1z0m7SrE/PDF/27tOFlK6sA0dc.pdf
dc.identifier.urihttp://edoc.rki.de/176904/2231
dc.description.abstractOscillometric blood pressure (BP) measurement devices are increasingly replacing standard mercury sphygmomanometers and generalizability of validation studies to other environments, for example, national survey environments, is assumed. We compared BP measurements according to two highly standardized German national survey BP protocols: a standard mercury sphygmomanometer and an oscillometric device, Datascope Accutorr Plus, each with specific manufacturer-provided cuffs and cuff-selection rules. A sample of 105 adults were subjected to alternate same-arm BP measurements according to the principles of the International Protocol revision 2010 for the validation of BP-measuring devices in adults of the European Society of Hypertension. In all, 315 BP measurement pairs were obtained. The mean systolic blood pressure (SBP) and diastolic blood pressure (DBP) were higher by the standard mercury old protocol and increased with BP, age, and pulse pressure, and were associated with the ratios of the cuff width to the arm circumference. The mean systolic difference (datascope new protocol−standard mercury old protocol) in participants with old protocol: for SBPeng
dc.language.isoeng
dc.publisherRobert Koch-Institut, Epidemiologie und Gesundheitsberichterstattung
dc.subjectHumanseng
dc.subjectFemaleeng
dc.subjectMaleeng
dc.subjectAdulteng
dc.subjectMiddle Agedeng
dc.subjectYoung Adulteng
dc.subjectHealth Surveyseng
dc.subjectBlood Pressure Determination/instrumentationeng
dc.subjectOscillometry/instrumentationeng
dc.subjectBlood Pressure/physiologyeng
dc.subjectBlood Pressure Determination/methodseng
dc.subjectBlood Pressure Monitors/standardseng
dc.subjectCalibrationeng
dc.subjectLinear Modelseng
dc.subjectSphygmomanometers/standardseng
dc.subject.ddc610 Medizin
dc.titleCalibration of blood pressure data after replacement of the standard mercury sphygmomanometer by an oscillometric device and concurrent change of cuffs
dc.typeperiodicalPart
dc.identifier.urnurn:nbn:de:0257-10042564
dc.identifier.doi10.1097/MBP.0000000000000081
dc.identifier.doihttp://dx.doi.org/10.25646/2156
local.edoc.container-titleBlood Pressure Monitoring
local.edoc.container-textThis is a non-final version of an article published in final form in Blood Pressure Monitoring 2015(20)1, pp. 39-42.
local.edoc.fp-subtypeArtikel
local.edoc.type-nameZeitschriftenartikel
local.edoc.container-typeperiodical
local.edoc.container-type-nameZeitschrift
local.edoc.container-urlhttp://journals.lww.com/bpmonitoring/Abstract/2015/02000/Calibration_of_blood_pressure_data_after.8.aspx
local.edoc.container-publisher-nameLippincott, Williams & Wilkins
local.edoc.container-volume20
local.edoc.container-issue1
local.edoc.container-year2015

Zur Kurzanzeige