Calibration of blood pressure data after replacement of the standard mercury sphygmomanometer by an oscillometric device and concurrent change of cuffs
dc.contributor.author | Neuhauser, Hannelore | |
dc.contributor.author | Ellert, Ute | |
dc.contributor.author | Thamm, Michael | |
dc.contributor.author | Adler, Carolin | |
dc.date.accessioned | 2018-05-07T18:46:09Z | |
dc.date.available | 2018-05-07T18:46:09Z | |
dc.date.created | 2015-12-23 | |
dc.date.issued | 2014-08-20 | none |
dc.identifier.other | http://edoc.rki.de/oa/articles/re1z0m7SrE/PDF/27tOFlK6sA0dc.pdf | |
dc.identifier.uri | http://edoc.rki.de/176904/2231 | |
dc.description.abstract | Oscillometric blood pressure (BP) measurement devices are increasingly replacing standard mercury sphygmomanometers and generalizability of validation studies to other environments, for example, national survey environments, is assumed. We compared BP measurements according to two highly standardized German national survey BP protocols: a standard mercury sphygmomanometer and an oscillometric device, Datascope Accutorr Plus, each with specific manufacturer-provided cuffs and cuff-selection rules. A sample of 105 adults were subjected to alternate same-arm BP measurements according to the principles of the International Protocol revision 2010 for the validation of BP-measuring devices in adults of the European Society of Hypertension. In all, 315 BP measurement pairs were obtained. The mean systolic blood pressure (SBP) and diastolic blood pressure (DBP) were higher by the standard mercury old protocol and increased with BP, age, and pulse pressure, and were associated with the ratios of the cuff width to the arm circumference. The mean systolic difference (datascope new protocol−standard mercury old protocol) in participants with old protocol: for SBP | eng |
dc.language.iso | eng | |
dc.publisher | Robert Koch-Institut, Epidemiologie und Gesundheitsberichterstattung | |
dc.subject | Humans | eng |
dc.subject | Female | eng |
dc.subject | Male | eng |
dc.subject | Adult | eng |
dc.subject | Middle Aged | eng |
dc.subject | Young Adult | eng |
dc.subject | Health Surveys | eng |
dc.subject | Blood Pressure Determination/instrumentation | eng |
dc.subject | Oscillometry/instrumentation | eng |
dc.subject | Blood Pressure/physiology | eng |
dc.subject | Blood Pressure Determination/methods | eng |
dc.subject | Blood Pressure Monitors/standards | eng |
dc.subject | Calibration | eng |
dc.subject | Linear Models | eng |
dc.subject | Sphygmomanometers/standards | eng |
dc.subject.ddc | 610 Medizin | |
dc.title | Calibration of blood pressure data after replacement of the standard mercury sphygmomanometer by an oscillometric device and concurrent change of cuffs | |
dc.type | periodicalPart | |
dc.identifier.urn | urn:nbn:de:0257-10042564 | |
dc.identifier.doi | 10.1097/MBP.0000000000000081 | |
dc.identifier.doi | http://dx.doi.org/10.25646/2156 | |
local.edoc.container-title | Blood Pressure Monitoring | |
local.edoc.container-text | This is a non-final version of an article published in final form in Blood Pressure Monitoring 2015(20)1, pp. 39-42. | |
local.edoc.fp-subtype | Artikel | |
local.edoc.type-name | Zeitschriftenartikel | |
local.edoc.container-type | periodical | |
local.edoc.container-type-name | Zeitschrift | |
local.edoc.container-url | http://journals.lww.com/bpmonitoring/Abstract/2015/02000/Calibration_of_blood_pressure_data_after.8.aspx | |
local.edoc.container-publisher-name | Lippincott, Williams & Wilkins | |
local.edoc.container-volume | 20 | |
local.edoc.container-issue | 1 | |
local.edoc.container-year | 2015 |